LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP (A21)/CO2 (A23) Module Replacement
296
8
6.
If present, connect CO2 module tube to CO2 inlet tube.
NOTE:
Older versions of the CO2 module have a Nafion inlet tube
which can be pinched if pushed back into the CO2 module
area. Keep the tube from the module stretched out while
connecting the inlet tube.
7.
If present, connect CO2 inlet cable (W28) to CO2 adapter cable (W30).
To install the A21 NIBP, A23 CO2, and SpO2 PCBs:
(Continued) 12 steps, (Page 5 of 6)
W28-W30
CO2 inlet cable
connector
CO2 inlet tube
CO2 hose from
module
Figure 8.92—CO2 module connections