LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Power/Therapy PCB Cable (W02) Replacement
325
8
Removing the Power/Therapy PCB Cable (W02)
To remove the power/therapy PCB cable from the rear case:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
PCB Assembly Replacement (p. 232)
3.
For orientation, locate power PCB-J8 on the rear case diagram. (The therapy PCB-J20 end
of the power/therapy PCB cable was disconnected as part of step 2.)
4.
Press the connector locking tab, and then disconnect the power/therapy PCB cable at J8.
Installing the Power/Therapy PCB Cable (W02)
To install the power/therapy PCB cable into the rear case:
1.
Connect the power/therapy PCB cable (
) to the power PCB (A03) at J8.
2.
Install the system/therapy PCB assembly as described in
Therapy (A04) PCB Assembly (p. 239)
3.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
P08
P20
to Power PCB (A03)
to A04 Therapy PCB
Figure 8.106—Power/therapy PCB
cable connections