LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
CO2 Inlet Connector Cable (W28) Replacement
351
8
Removing the CO2 Inlet Connector Cable (W28)
To remove the CO2 inlet connector cable (
) from the parameter bezel (rear case—refer to
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
Removing the System (A01)/Therapy (A04) PCB Assembly (p. 233)
3.
Remove the parameter bezel as described in
Removing the Parameter Bezel (p. 311)
.
4.
Remove the four screws (
) securing the CO2 connector adapter (
5.
Remove the CO2 connector adapter and, from the underside of the bezel, remove the CO2 connector retainer (
and seal (
). Discard the seal and screws.
6.
Disengage the CO2 inlet connector cable from the CO2 connector retainer by spreading the arms of the retainer and sliding out
the CO2 inlet connector body (see