Replacement Procedures
NIBP (A21)/CO2 (A23) Module Replacement
297
8
8.
If present, connect the exhaust tube to the rear case outlet connector.
9.
Install the OEM PCB as described in
Installing the OEM PCB (A06) (p.
.
10. Install the transfer relay assembly as described in
.
11. Install the system/therapy PCB assembly as described in
Therapy (A04) PCB Assembly Replacement (p. 232)
.
12. Reassemble the case as described in
Reassembling the Case (p. 184)
.
To install the A21 NIBP, A23 CO2, and SpO2 PCBs:
(Continued) 12 steps, (Page 6 of 6)
Figure 8.93—CO2 exhaust tube connection