Replacement Procedures
System Connector Cable (W08) and Auxiliary Connector Cable
(W09) Replacement
333
8
Installing the System Connector Cable (W08) or Auxiliary Power Cable (W09)
Refer to
To install the system connector cable or auxiliary power cable into the rear case:
NOTE:
When installing a new system connector cable (
) or auxiliary power cable (V2 part
), use the
Internal Hardware Repair Kit (MIN 3305431-015) (p. 507)
External Hardware Repair Kit (MIN 3305431-016) (p.
1.
If removed, install the system connector cable in hole above the auxiliary power cable connector.
2.
) over wire connector of the system cable.
3.
Thread seal plate (
) over wire connector of the system cable (torque to 10 in-lb
using a 3/4” modified deep socket).
4.
If removed, install auxiliary power cable in the lower hole provided in rear case.
5.
) over wire connector of the auxiliary power cable.
6.
Thread seal plate (
) (flat side down) and nut (
) over wire connector of the auxiliary power cable (torque to
10 in-lb using a 3/4” modified deep socket).
7.
Install the power PCB as described in
Installing the Power PCB (A03) (p. 257)
8.
Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
9.
Install the system/therapy PCB assembly as described in
Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239)
10. Reassemble the case as described in
Reassembling the Case (p. 184)
.
CAUTION
POSSIBLE MOISTURE LEAKAGE
When installing the system connector cable, use a
new O-ring seal to help prevent ingress of fluids.