Programming Site Data
1-57
12/1/99
The COMM Sub-Menu
The Model GF868 flowmeter can transmit stored data and displayed
readings to a remote ANSI terminal or a personal computer by
connecting the meter’s RS232 interface to the serial port of the PC. In
addition, the Model GF868 can receive and execute remote
commands via this link.
Use the
COMM
sub-menu to set the RS232 communications port
parameters and to enter a network identification number. A network
identification number is required to use the Panametrics Flowmeter
Data Manager software. While following the programming
instructions, refer to the menu map in Figure A-3 on page A-3 in
Appendix A, Menu Maps. Enter the
COMM
sub-menu by pressing
[
→]
and
[F3]
at the initial User Program screen and complete the
following steps:
The available rates at the above prompt are 300, 600, 1200, 2400,
4800, 9600 and 19200 baud.
The five available responses for the
UART bits
prompt are listed in
Table 1-18 below.
Global COMM PROGRAM
É
Press the [
→]
until the desired
baud rate appears on the option
bar and then press the
appropriate
[Fx]
function key to
select it.
Global PROGRAM
Comm port
BAUD RATE
current value appears here
4800
9600
19200
Global COMM PROGRAM
É
Press the [
→]
until the desired
setting appears on the option bar
and then press the appropriate
[Fx]
function key to select it.
BAUD RATE
current value appears here
UART bits
current setting appears here
8,no
8,odd
8even
7,no
Table 1-18: UART Bits Options
Option Bar
# Data Bits
# Stop Bits
Parity
8,no
8
0
None
8,odd
8
0
Odd
8even
8
0
Even
7,odd
7
1
Odd
7even
7
1
Even
Summary of Contents for GF868
Page 1: ...Model GF868 Programming Manual Two Channel...
Page 6: ...Chapter 1...
Page 71: ...Chapter 2...
Page 92: ...Chapter 3...
Page 113: ...Chapter 4...
Page 131: ...Chapter 5...
Page 138: ...12 1 99 5 6 Clearing Data...
Page 139: ...Chapter 6...
Page 150: ...Appendix A...
Page 159: ...Appendix B...
Page 160: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2...