Programming Site Data
1-17
12/1/99
The PIPE Sub-Menu
(cont.)
If the pipe wall thickness is not known and can not be conveniently
measured, look up the value in a table of standard pipe size data or
use the Model GF868’s on-line Help Menu.
Note:
Panametrics has calculated both the transducer signal path
length
(P)
and the transducer signal axial length
(L)
, based on
the exact transducer configuration used for the application.
These values are engraved on the flowcell and/or are included
in the documentation supplied with the meter.
Chan 1 PIPE PROG
Press
[F1]
or
[F2]
to select the
desired units. Then, enter the
path length of the ultrasonic
signal and press
[ENT]
. (The
meter will only accept values
from 1/8 to 900 in.)
PIPE WALL
current value appears here
PATH LENGTH P
current value appears here
inch
feet
Chan 1 PIPE PROG
Press
[F1]
or
[F2]
to select the
desired units. Then, enter the
axial length of the ultrasonic
signal and press
[ENT]
. (The
meter will only accept values
from 1/8 to 900 in.)
PATH LENGTH P
current value appears here
AXIAL LENGTH L
current value appears here
inch
feet
Chan 1 PIPE PROG
Enter a value for the flow
calibration factor and press
[ENT]. The default value is 1.00.
(The meter will only accept
values from 0.50000 to
2.00000.)
AXIAL LENGTH L
current value appears here
CALIBRATION FACTOR
current setting appears here
Chan 1 PIPE PROG
Press
[F1]
to disable this
function, or press
[F2]
to enable
multiple K-factors.
CALIBRATION FACTOR
previous entry appears here
Multi K-factors
current status appears here
OFF
ON
Summary of Contents for GF868
Page 1: ...Model GF868 Programming Manual Two Channel...
Page 6: ...Chapter 1...
Page 71: ...Chapter 2...
Page 92: ...Chapter 3...
Page 113: ...Chapter 4...
Page 131: ...Chapter 5...
Page 138: ...12 1 99 5 6 Clearing Data...
Page 139: ...Chapter 6...
Page 150: ...Appendix A...
Page 159: ...Appendix B...
Page 160: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2...