12/1/99
1-40
Programming Site Data
The ERROR Option
This menu option permits programming of the manner in which the
Model GF868’s outputs respond during an error condition. See
Chapter 2, Error Codes and Screen Messages, of the Service Manual
for a discussion of the built-in error codes.
Note:
The error responses listed in Table 1-7 above apply only to the
CH1/CH2
measurement mode and to the
SUM
/
DIF
measurement modes, if either or both channels are in error.
Error handling in the
AVE
measurement mode is set below.
The
2-path error handling
option is intended for applications where
two sets of transducers are installed in the same location in the same
pipe to improve accuracy and the meter is operated in
AVE
mode.
With this function enabled, the Model GF868 performs error handling
only if both channels are in error. If this function is disabled, error
handling occurs when either channel goes into error.
Global Error Action
Press
[F1]-[F4]
to select the
desired error response for the
CH1/CH2
,
SUM
or
DIF
modes,
as described in Table 1-7 below.
Global I/O
Error handling
ERROR HANDLING
last selection appears here
HOLD
LOW
HIGH
HHIGH
Table 1-7: Standard Error Response Options
Option
Display Response
Totalizer Response
HOLD
Holds the last “good”
reading
Holds the last “good”
reading and continues to
totalize, based on that
reading
LOW
Forces the output to the
low set point
Stops totalizing
HIGH
Forces the output to the
high set point
Stops totalizing
HHIGH
Forces the output
≈
10%
above the high set point
Stops totalizing
Global Error Action
Press
[F1]
to disable 2-path error
handling, or press
[F2]
to enable
2-path error handling. The User
Program automatically returns to
the
Global I/O
prompt.
ERROR HANDLING
previous selection appears here
2PATH ERROR HANDLING
last selection appears here
NO
YES
Summary of Contents for GF868
Page 1: ...Model GF868 Programming Manual Two Channel...
Page 6: ...Chapter 1...
Page 71: ...Chapter 2...
Page 92: ...Chapter 3...
Page 113: ...Chapter 4...
Page 131: ...Chapter 5...
Page 138: ...12 1 99 5 6 Clearing Data...
Page 139: ...Chapter 6...
Page 150: ...Appendix A...
Page 159: ...Appendix B...
Page 160: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2...