Programming Site Data
1-27
12/1/99
The SIGNL Option(cont.)
The percentage of peak used to calculate the transit times and Delta T
is specified at this prompt. The default value for this parameter is
50% and values from 1 to 100% are acceptable.
Both the upstream and downstream transducers transmit ultrasonic
pulses in bursts, which consist of a series of transmit pulses.
XMIT
SAMPLE SIZE
determines how many bursts are sent in one direction
before sending in the other direction. The default value for this
parameter is 8 and values of 2, 4, 8, 16 and 32 are acceptable.
If the burst mode is set to Skan/Measure (S/M), the meter switches
from Skan to Measure Mode when Delta T is less than the
M>S_Switch
value.The default value for this parameter is 50
µ
sec and
values from 0 to 250
µ
sec are acceptable.
Chan 1 SIGNAL SETUP
Press
[ENT]
to accept the current
% of Peak
value or enter a new
value and press
[ENT]
.
Skan T OFFSET
current value appears here
% of Peak
current value appears here
Chan 1 SIGNAL SETUP
É
Use the [
→]
and [F1]-[F4] keys
to select one of the preset values
for the
XMIT SAMPLE SIZE
from the option bar.
% of Peak
current value appears here
XMIT SAMPLE SIZE
current value appears here
2
4
8
16
Chan 1 SIGNAL SETUP
Press
[ENT]
to accept the current
M>S_Switch
value or enter a
new value and press
[ENT]
.(DO
NOT change this value unless
advised to do so by the factory).
XMIT SAMPLE SIZE
current value appears here
M>S_Switch
current value appears here
Summary of Contents for GF868
Page 1: ...Model GF868 Programming Manual Two Channel...
Page 6: ...Chapter 1...
Page 71: ...Chapter 2...
Page 92: ...Chapter 3...
Page 113: ...Chapter 4...
Page 131: ...Chapter 5...
Page 138: ...12 1 99 5 6 Clearing Data...
Page 139: ...Chapter 6...
Page 150: ...Appendix A...
Page 159: ...Appendix B...
Page 160: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2...