Programming Site Data
1-51
12/1/99
Option Card Totalizer/
Frequency Outputs (cont.)
Note:
The measurement units that appear in these prompts are those
selected in the
GLOBL-SYSTM
menu earlier in this chapter.
Skip over section B and rejoin the programming sequence.
B: Programming a Totalizer Output
This type of output issues one pulse per selected volume of flow. The
meter produces a pulse each time the programmed amount of flow
passes through the pipe.
Note:
The measurement units that appear in these prompts are those
selected in the
GLOBL-SYSTM
menu earlier in this chapter.
Global Pulse Output
Enter a value for the low end of
the frequency output range and
press
[ENT]
.
Output measurement
last selection appears here
BASE
current value appears here
Global Pulse Output
Enter a value for the high end of
the frequency output range and
press
[ENT]
.
BASE
current value appears here
FULL
current value appears here
Global Pulse Output
Enter a value between 1 and
10,000 for the frequency at full
scale and press
[ENT]
.
FULL
current value appears here
FULL SCALE FREQUENCY
current value appears here
Global Pulse Output
Press
[F1]
to totalize the forward
flow or
[F2]
to totalize the
reverse flow. See Table 1-17 on
page 1-52 for a description of the
available options.
Slot x Pulse Out A
Totalizer
Entity to totalize
last selection appears here
+TOTL
-TOTL
+MASS -MASS
Summary of Contents for GF868
Page 1: ...Model GF868 Programming Manual Two Channel...
Page 6: ...Chapter 1...
Page 71: ...Chapter 2...
Page 92: ...Chapter 3...
Page 113: ...Chapter 4...
Page 131: ...Chapter 5...
Page 138: ...12 1 99 5 6 Clearing Data...
Page 139: ...Chapter 6...
Page 150: ...Appendix A...
Page 159: ...Appendix B...
Page 160: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2...