AFX SERIES® OPERATION MANUAL
SECTION 8: REMOTE PROGRAMMING
Entire Contents Copyright
2016 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 155 of 274
8.6.1.2
Transient Segment Output Parameter List Table
The following table details the available parameters for the transient segment program
definitions.
Parameters
Description
SEGment,<n>
Transient segment n of the selected program to be edited. Also the
starting segment when querying multiple segments (see NSEGS)
n = <1> to <100>
FSEG,<n>
sets objective Frequency of selected segment
n = <:SOUR:FREQ:LIM:MIN> to <:SOUR:FREQ:LIM:MAX> Hz
VSEG,<n>
sets objective voltage ФA, ФB, ФC of selected segment
n = <0> to <300 x XFMRRATIO> Volts
VSEG1,<n>
sets objective voltage ФA of selected segment
n = <0> to <300 x XFMRRATIO> Volts
VSEG2,<n>
sets objective voltage ФB of se
lected segment
n = <0> to <300 x XFMRRATIO> Volts
VSEG3,<n>
sets objective voltage ФC of selected segment
n = <0> to <300 x XFMRRATIO> Volts
WFSEG,<n>
sets Waveform ФA, ФB, ФC of selected segment
n = <1> to <16>
WFSEG1,<n>
sets Waveform ФA of selected
segment
n = <1> to <16>
WFSEG2,<n>
sets Waveform ФB of selected segment
n = <1> to <16>
WFSEG3,<n>
sets Waveform ФC of selected segment
n = <1> to <16>
TSEG,<n>
sets execution time (to reach objective Voltage and Frequency) of
selected segment
n = <0> or <0.0002> to <300> seconds. Setting n to <0> selects
cycle-based transient operation, i.e., each segment time is equal to
the period of 1 cycle (1/FREQ)
LAST
sets selected segment to be the last segment of selected transient
Parameter
Description
NSEGS,<n>
Command: Specifies the number of Transient segments (all
parameters) to return in a subsequent :PROG:DEFine? query. If
NSEGS is not specified, 1 segment is returned by :PROG:DEFine?
EXAMPLE:
:PROG:DEF SEG,3,NSEGS,4;PROG:DEF? will return 4
transient segments, starting with Segment 3.
NOTE:
The value will revert back to 1 after each query.
Query: The NSEGS parameter of a :PROG:DEF? query returns total
number of Transient segments defined in the selected program