AFX SERIES® OPERATION MANUAL
SECTION 6: FRONT PANEL OPERATION
Entire Contents Copyright
2016 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 105 of 274
Update Phase
Sets the phase angle at which output voltage and frequency
changes will take place on phase A. This setting applies to
both steady state output changes and to the start of a
transient program execution. It also applies to the OUTPUT
ENABLE and DISABLE function key on the front panel.
Changes on phases B and C will take place at the same
moment in time but at phase angles that are shifted from
phase A by the phase angles programmed for phase B and
C.
Allow Output Enable at Power On
This mode if set, causes the output to turn on at
power up.
Enable Current Overload
Current Overload mode allows the power source to provide
more than the nominal max. RMS current (around 30
percent more) for a short period of time (up to 2 seconds).
When enabled, the user can set the CURR:LIM 30% higher
than whit this mode off. For example in a stand-alone
3150AFX unit in three phase mode (FORM 3), RMS output
current can be up to 55A when the continuous output limit
is 41.7A. A warning message will be display notifying the
user this mode is being enabled.
Access to other utility screens is available from the UNIT CONFIGURATION screen through
the soft keys. The table below lists the available soft keys on the UNIT CONFIGURATION
screen.
SOFT KEY
Description
User Limits (SK1)
Displays SETUP MENU screen. Refer to section 6.7.2
Ramp & Slew (SK3)
Displays SLEW RATE MENU screen. Refer to section 6.7.3
Program Memory (SK2)
Displays PROGRAM MEMORY screen. Refer to section 6.7.4
CSC Config. (SK4)
Display CSC setting screen
Table 6-15: Available CONFIGURATION screen soft keys
Summary of Contents for 3120AFX
Page 273: ......