Query Format
[SOURce:]VOLTage:DC#:SATuration[:LIMit]?
Description
Returns maximum DC voltage at which saturation of the output
inverter will occur. Saturation limits are determined in the maximum
peak voltage capability of the AC+DC components at the output of the
power source. For the AFX Series®, these limits are -425 and +425V.
The VOLT:DC1:SAT? returns the max. DC level that may be
programmed based on the programmed AC wave shape and RMS level.
Maximum allowable values are:
Maximum positive peak voltage:
V_DC = V_peak_max - V_AC_peak_max
Minimum negative peak voltage:
V_DC = V_peak_min + V_AC_peak_min
Examples
V_AC = 300Vrms sine wave
V_DC = 425 - 1.41 *
300
(
V_DC_sat = 0
)
V_AC = 120Vrms sine wave
V_DC = 425 - 1.41 *
120
(
V_DC_sat = 255
)
V_AC = 0Vrms
V_DC = 425 –
0
(
V_DC_sat = 425
)
Returned Data Format
<nr2>
Query Example
VOLT:DC1:SAT?
-225, + 225
Command Syntax
[SOURce:]VOLTage:DC:INITial
Description
This command is equivalent to the “[SOURce:]INITial <n>” command
but applies only to the DC1, DC2 and DC3 parameters.
VOLT:DC:INIT? = ON Set points of the unit at power on will be the last set after power off.
The set points affected by this command are:
- Voltage DC1
- Voltage DC2
- Voltage DC3
VOLT:DC:INIT? = OFF Set points of the unit at power on will have a default value of:
- Voltage DC1 = 0V
- Voltage DC2 = 0V
- Voltage DC3 = 0V
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<b>
Example
VOLT:DC:INIT OFF
Query Format
[SOURce:]VOLTage:DC:INITial?
Returned Data Format
<b>
Query Example
VOLT:DC:INIT?
0