AFX SERIES® OPERATION MANUAL
SECTION 8: REMOTE PROGRAMMING
Entire Contents Copyright
2016 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 175 of 274
Command Syntax
[SOURce:]VOLTage[:AC]#
Description
Sets AC voltage for phase #. If # is omitted, sets all available phases to
value specified.
Parameters
Range 0.0000 - 300.0000
Parameter Format
<nr2>
Example
VOLT:AC1 100.00
Query Format
[SOURce:]VOLTage[:AC]#?
Description
Returns voltage setting for specified phase. If phase is omitted, returns
voltage setting for phase A.
Returned Data Format
<nr2>
Query Example
VOLT:AC1?
100.0000
Command Syntax
[SOURce:]VOLTage[:AC]#:SLEW
Description
Sets AC voltage slew rate for phase # in Volts per msec. If # is omitted,
sets phase A slew rate to value specified.
Parameters
Range 0.01 – 300.0
Parameter Format
<nr2> and <nr3>
Example
VOLT:AC1:SLEW 300.00
Query Format
[SOURce:]VOLTage[:AC]#:SLEW?
Description
Returns voltage slew rate setting for specified phase. If phase is
omitted, returns voltage setting for phase A.
Returned Data Format
<nr2>
Query Example
VOLT:AC1:SLEW?
300.0000
Command Syntax
[SOURce:]VOLTage[:AC]:LIMit:MINimum
Description
Sets low user limit for AC voltage programming. User limits must fall
within actual hardware limits of the power source. Also, MIN limit must
be less than MAX limit.
Parameters
0.000 – 300.000
Parameter Format
<nr2>
Example
VOLT:LIM:MIN 20.0
Query Format
[SOURce:]VOLTage[:AC]#:LIMit:MINimum?
Returned Data Format
<nr2>
Query Example
VOLT:LIM:MIN?
20.0000