Parameter
DDD
VVI
AAI
VDD
Unit
Basic Rate
72
72
72
72
ppm
Atrial Overdrive Stimulation
Start Rate
240
–
240
–
ppm
Ventricular Rapid Pacing
Start Rate
180
180
–
180
ppm
Atrial Amplitude
5
–
5
–
V
Atrial Sensitivity
1
–
1
0.5
mV
Atrial Pulse Duration
1
–
1
–
ms
Ventricular Amplitude
8
8
–
8
V
Ventricular Pulse Duration
0.75
0.75
–
0.75
ms
Ventricular Sensitivity
2
2
–
2
mV
A-V Delay
AUTO
i.e. 180
–
–
AUTO
i.e. 180
ms
PVARP
AUTO
i.e. 280
–
–
AUTO
i.e. 280
ms
MTR
AUTO
i.e. 106
–
–
AUTO
i.e. 106
ms
Automatic Sensing
OFF
OFF
OFF
OFF
–
Atrial Trigger Option
OFF
–
OFF
–
–
Table 38: Manufacturer’s Default Programs