PACE 203H – Instructions for Use
62 / 139
5I-17-016X-B-20
Pulse Duration-Menu
Key No.
A
↑
1
1.00 ms
↓
2
V
↑
3
0.75 ms
↓
4
↑
_
5
Table 46: Pulse Duration Menu
The pulse duration can be changed with
the softkey labeled
↑
or the softkey la-
beled
↓
in increments / decrements of
0.05 ms between 0.05 ms and 1.50 ms,
respectively, as shown in Table 47.
The parameter change appears immedi-
ately in corresponding line of the lower
display and will become effective with
the next intervention interval.
Parameter
Range (Step Width)
Unit
Atrial Pulse
Duration
0.05 (0.05) 1.50
ms
Ventricular
Pulse Duration
0.05 (0.05) 1.50
ms
Table 47: Ranges and step widths (in
parenthesis) for manual setting of Pulse
Durations
9.12.4 Adjustment of Atrial (ARP) and
Ventricular (VRP) Refractory Pe-
riod
The atrial refractory period (ARP) and
the ventricular refractory period (VRP;
Firmware Version 1.24 and up) of this
pacemaker can be changed.
By pressing the softkey labeled
ARP & VRP
in the Parameters/Options Menu (Table
41 on page 60), the ARP & VRP Menu
appears as shown in Table 48.
The ARP & VRP Menu can be used to
change the setting for ARP and VRP.
ARP & VRP Menu
Key No.
ARP
↑
1
250 ms
↓
2
VRP
↑
3
250 ms
↓
4
↑
_
5
Table 48: ARP & VRP Menu
The ARP can be changed with the
softkey labeled
ARP
↑
or the softkey
labeled
ARP
↓
in increments / decre-
ments of 10 ms between 250 ms and
400 ms, respectively, as shown in Table
49. The manufacturer's defaults setting
is 250 ms (refer to section 9.11.4 on
page 56).
Summary of Contents for PACE 203H
Page 2: ......
Page 8: ......
Page 16: ......
Page 20: ......
Page 102: ......
Page 106: ......
Page 108: ......
Page 116: ......
Page 118: ......
Page 124: ...PACE 203H Instructions for Use 124 139 5I 17 016X B 20 Figure 18 Map of Menu Tree Part 2 ...
Page 130: ......
Page 132: ......
Page 136: ......
Page 138: ......