PACE 203H – Instructions for Use
88 / 139
5I-17-016X-B-20
Abnormal Settings
Reason for
Preventing
Mode
Action
a) Basic Rate
>
MTR
Basic Rate must not
be larger than MTR
DDD
VDD
DAI
VAT
DAT
Transient warning message and
prevent abnormal setting
b) AVD + PVARP + 20ms
>
LRI
To guarantee mini-
mum atrial sensing
phase
DDD
VDD
DAI
VAT
DAT
Transient warning message and
prevent abnormal setting
c) AVD + 60ms
>
LRI
To guarantee mini-
mum VA-delay
D00
DVI
Transient warning message and
prevent abnormal setting
d) PVD + PVARP
≥
URI
(with URI = 1/MTR)
To guarantee
Wenckebach-
behavior
DDD
VDD
DAI
VAT
DAT
Transient warning message and
prevent abnormal setting
e) ARP + 20 ms > LRI
To guarantee mini-
mum atrial sensing
phase
AAI
AAT
Transient warning message and
prevent abnormal settings
Table 60: Abnormal Settings
9.22.7.1 Possible Conflicts
Conflicts can occur in the process of
changing parameter settings. In the tab-
les below, all possible cases are listed
together with the transient conflict warn-
ing message, which will be displayed on
the lower display. In case of a conflict,
the parameter to be changed is limited to
a value still permissible.
Summary of Contents for PACE 203H
Page 2: ......
Page 8: ......
Page 16: ......
Page 20: ......
Page 102: ......
Page 106: ......
Page 108: ......
Page 116: ......
Page 118: ......
Page 124: ...PACE 203H Instructions for Use 124 139 5I 17 016X B 20 Figure 18 Map of Menu Tree Part 2 ...
Page 130: ......
Page 132: ......
Page 136: ......
Page 138: ......