PACE 203H – Instructions for Use
58 / 139
5I-17-016X-B-20
9.11.5 Changing the Turn-On
Program
By pressing the softkey
Save as Turn-on
in the Standard Menu (Table 35), the
actual setting will become the permanent
Turn-On program when the question
Overwrite Turn-on program?
is answered with "YES" (Table 39).
Turn-on Saving
Confirmation
Key No.
Overwrite
Turn-on
program?
1
2
YES
3
NO
4
↑
_
5
Table 39: Turn-on Saving Confirmation
After storing the user defined Turn-on
Program, PACE 203H abandons the
Turn-On Menu and returns to the Mode
Menu.
If answered with "NO", PACE 203H
returns to the (previous) Standard Menu.
Pressing the
↑
_
key forces PACE 203H
to revert to the Mode Menu.
Note
: Upon the manufacturer’s shipping
of PACE 203H, the Turn-On Program
parameters are set according to Table
40.
Parameter
Value
Unit
Mode
DDD
–
Basic Rate
60
ppm
Start Rate Atrial Over-
drive Stimulation
240
ppm
Start Rate Ventricular
Rapid Pacing
180
ppm
Atrial Amplitude
5
V
Atrial Sensitivity
0.5
mV
Atrial Pulse Duration
1
ms
Ventricular Amplitude
8
V
Ventricular Pulse
Duration
0.75
ms
Ventricular Sensitivity
1
mV
A-V Delay
AUTO
i.e. 200
ms
PVARP
AUTO
i.e. 280
ms
MTR
AUTO
i.e. 100
ppm
Automatic Sensing
OFF
–
Atrial Trigger Option
OFF
–
Table 40: Manufacturer’s Turn-on
Program
Summary of Contents for PACE 203H
Page 2: ......
Page 8: ......
Page 16: ......
Page 20: ......
Page 102: ......
Page 106: ......
Page 108: ......
Page 116: ......
Page 118: ......
Page 124: ...PACE 203H Instructions for Use 124 139 5I 17 016X B 20 Figure 18 Map of Menu Tree Part 2 ...
Page 130: ......
Page 132: ......
Page 136: ......
Page 138: ......