PACE 203H – Instructions for Use
112 / 139
5I-17-016X-B-20
Parameter
Basic rate 80 ppm,
Amplitude 18 V (atrium and ventricle),
Pulse duration 1.5 ms (atrium and ventricle),
AVD 170 ms (auto)
Standard setting:
Standard parameter set for each primary
mode,
plus one turn-on parameter set
Acoustic signaling:
Different for stimulation, sensing and
warnings; stimulation and sensing indication
can be turned on or off
Safety
Interference detection frequency:
> 4.5 Hz ± 5 % (> 270 ppm)
Interference rate:
Basic Rate + 10 ppm ± 2 %
(max. MTR and max. 220 ppm)
Defibrillation protection:
According to IEC 60601-2-31
Runaway protection:
238 ppm ± 3 ppm
Lead surveillance:
Message in case of short circuit or inter-
ruption in the stimulation circuit (for a set
stimulation amplitude equal or higher 2.0 V
and a set pulse duration equal or higher
0.15 ms)
Battery
Battery:
9 Volt (identification per IEC 86: 6LR61)
Recommended types:
Duracell
®
Procell
®
MN1604
Ultralife
®
Lithium Power Cell
®
Life span of recommended battery
minimal 10 days (VVI, standard parameter)