PACE 203H – Instructions for Use
44 / 139
5I-17-016X-B-20
Auto Sense Menu
Key No.
1
Atrial
Auto Sense
2
Ventricular
Auto Sense
3
Initiate
(Press & Hold)
4
↑
_
5
Table 22: Auto Sense Menu
(channels selected, initialization request)
When the softkey labeled
Initiate
is
pressed, PACE 203H stops pacing as
long as this key is held down. However,
PACE 203H continues to sense and
attempts to detect a regular intrinsic
activation of the patient’s heart. While
doing this, the lower display appears as
shown in Table 23. A flashing heart
symbol indicates each detected event.
Furthermore, the acoustic indicator will
automatically be enabled temporarily, if it
is not already switched on. Because
PACE 203H goes through an initializa-
tion algorithm, individual intrinsic events
may or may not be detected.
Auto Sense Menu
Key No.
1
Searching
A-Sense
♥
2
Searching
V-Sense
♥
3
hold...
4
5
Table 23: Auto Sense Menu (Searching)
If PACE 203H has found an intrinsic
heart rhythm, it will display the sensed
EGM amplitude as shown in Table 24.
Auto Sense Menu
Key No.
Sensed:
1
A = 2.3 mV
2
V = 8.0 mV
3
Finished
(release key)
4
5
Table 24: Auto Sense Menu
(intrinsic rhythm found)
The softkey labeled
Initiate
can now be
released. An appropriate sensitivity val-
ue will be automatically selected and set
which will be visible in the upper display.
Furthermore the sign
AUTO
will be
Summary of Contents for PACE 203H
Page 2: ......
Page 8: ......
Page 16: ......
Page 20: ......
Page 102: ......
Page 106: ......
Page 108: ......
Page 116: ......
Page 118: ......
Page 124: ...PACE 203H Instructions for Use 124 139 5I 17 016X B 20 Figure 18 Map of Menu Tree Part 2 ...
Page 130: ......
Page 132: ......
Page 136: ......
Page 138: ......