PACE 203H – Instructions for Use
5I-17-016X-B-20
19 / 139
25. All automatic settings provided by
PACE 203H are only a guide to as-
sist the user in finding the appropri-
ate settings. It is the responsibility of
the operator to determine if these
settings are correct.
26. All battery powered devices can lose
their function due to normal dis-
charge of the battery. It must also be
considered that the pacemaker can
stop operating as a result of an un-
predictable component failure. The
patient should not be left unattended
by the medical staff and should al-
ways be supervised by a monitoring
system.
27. If the battery must be replaced while
the device is being used on a patient,
ensure that PACE 203H was turned
on and operated for at least one
hour, allowing sufficient time to re-
charge the internal capacitors which
will maintain device operation for up
to 30 s while the battery is being re-
placed. If it is planned to replace the
battery during use of the device on a
patient, then prior use PACE 203H
must be checked for sufficient battery
backup time.
28. Use only 9 V primary batteries for
PACE 203H. Re-chargeable batter-
ies (accumulators) should not be
used (possibly low capacity and un-
stable charge condition may cause a
malfunction of the pacemaker).
29. When used as intended, PACE 203H
does not contain any parts which are
subject to wear and tear.
Summary of Contents for PACE 203H
Page 2: ......
Page 8: ......
Page 16: ......
Page 20: ......
Page 102: ......
Page 106: ......
Page 108: ......
Page 116: ......
Page 118: ......
Page 124: ...PACE 203H Instructions for Use 124 139 5I 17 016X B 20 Figure 18 Map of Menu Tree Part 2 ...
Page 130: ......
Page 132: ......
Page 136: ......
Page 138: ......