5I-17-016X-B-20
Parameter
Measurement conditions:
Environment temperature: 20 ± 2°C
Voltage supply: 9 ± 0.5 V
Load resistance: 500
Ω
± 1%
Test impulse: as specified in
ISO 14702-8 (Triangle 2 ms / 13 ms)
Pacing modes:
primary:
DDD, VVI, AAI, VDD
supplementary:
D00, V00, A00, DVI, DAI, VAT, AAT,
DDD+AT, DAT
Basic Rate:
30 … 220 ppm ± 2 %
Maximum tracking rate (MTR):
80 … 230 ppm ± 2 %
Atrial overdrive stimulation:
70 … 1000 ppm ± 5 %
Ventricular rapid pacing
(Firmware version 1.24 and up):
70 … 220 ppm ± 2 %
Output pulse:
Cathodic, biphasic, asymmetric,
capacitive coupled, passive discharge
Nominal pulse amplitude (500
Ω
):
0.1…18 V ± 10 % ± 0.05V
Nominal pulse amplitude
(200
Ω
… 2000
Ω
):
0.1…18 V ± 20 % ± 0.05V
Pulse duration:
0.05 ms ... 1.50 ms ± 10 % ± 0.01 ms
Atrial sensitivity:
0.2 … 20 mV ± 20 % ± 0.1 mV,
∞
Ventricular sensitivity:
1.0 … 20 mV ± 20 % ± 0.1 mV,
∞
Input impedance:
22 k
Ω
± 20 %
Output impedance:
< 10
Ω
Load impedance range:
200
Ω
…2000
Ω
A-V Delay (AVD):
5 ms ... 400 ms ± 5 % ± 2 ms
(minimum; 30 ms when atrial Auto Sense is
activated)
P-V delay (PVD):
Determined by A-V Delay:
PVD = A-V Delay - 30 ms ± 5 % ± 2 ms
Summary of Contents for PACE 203H
Page 2: ......
Page 8: ......
Page 16: ......
Page 20: ......
Page 102: ......
Page 106: ......
Page 108: ......
Page 116: ......
Page 118: ......
Page 124: ...PACE 203H Instructions for Use 124 139 5I 17 016X B 20 Figure 18 Map of Menu Tree Part 2 ...
Page 130: ......
Page 132: ......
Page 136: ......
Page 138: ......