1-24
Auto Focus Multi Code Reader V430-series User’s Manual (Z392)
Calibrate
Step 7 — Calibrate
V430-series settings can be adjusted automatically in
ESP
for optimum symbol decod-
ing performance.
During the calibration routine, the reader will flash its amber Read Rate percent LEDs and
red illumination LEDs while searching camera settings and determining the best configuration
for decoding symbol data. Upon successful completion of this routine, a green LED pattern
will flash brightly and illuminate the symbol.
Calibrate by ESP
• Click the
Calibrate
button.
• The reader will search camera settings to determine the best configuration for decoding
symbol data.
A successful calibration will display a green frame around the symbol, and the following
message will appear:
Uploading all reader parameters.
After a moment the symbol data
will be presented in the field below the image display window.
Note:
If you choose to calibrate the reader by sending a
<@CAL>
command, a decode is
also required for the calibration process to be completed successfully.