Configuration Aspects for Programmers
Appendix B
B-8
ChartScan User’s Manual
CSN14/TC/P
Configuring Thermocouple Channels
Begin configuring input channels by first specifying the channel type. When the CSN14/TC/P signal
conditioning module is used, 16 differential thermocouple inputs are provided. In configuring thermocouple
(T/C) channels, all that is required is to assign each channel to a particular T/C type. Support for B, E, J, K, R,
S, T and N (14 and 28 gauge) thermocouple types are standard. The 100 mV, ±1 V, ±5 V, ±10 V, volts types
with and without range-error detection are also supported. Channel type can be defined using the Configure
Channel (
C
) command for a single channel or a range of channels. Channels not included in the channel scan do
not have to be configured. The following table highlights channel types for normal use of the CSN14/TC/P
signal conditioning card.
The
type
parameter is what distinguishes the channel as a thermocouple channel. The channel configuration
cannot be changed while the trigger is armed or an acquisition is taking place.
In addition to channel type, each channel in the scan channel can optionally have a high and low alarm level and
hysteresis associated with it. ChartScan alarms allow the unit to produce an internal event when the value of a
channel is outside of desired limits. These alarm events can be used as trigger or stop events, or as a stimulus
for TTL-level signals on ChartScan back panel.
The following example demonstrates how to assign a J type thermocouple to channels on a CSN14/TC/P signal
conditioning module plugged into ChartScan.
PRINT#1 “OUTPUT07; C 1-16, 1 X” ‘Continuous channels 1-16 with JTC
Note:
Additional information can be found under the command reference for
C
in
Appendix A.
CSN14/LV/ (T, B, or S)
Configuring Low Volts Channels
When the CSN14/LV/ (T, B, or S) signal conditioning module is used, ChartScan measures voltages providing a
programmable range per channel. To configure the channels on a CSN14/LV/ (T, B, or S), use the Configure
Channel (
C
) command to assign a range to each channel on the signal conditioning module. Channels not
included in the channel scan do not have to be configured.
The following table highlights channel types for the CSN14/LV/ (T, B, and S) signal conditioning card.
Additional information can be found under the command reference for
C
in
Appendix A
.
The
type
parameter is what distinguishes the channel as a low volts channel. The channel configuration cannot
be changed while the trigger is armed or an acquisition is taking place.
Note:
Additional information can be found under the command reference for
C
in
Appendix A
.
Type
T/C
Type
T/C
Type
T/C
1
J
4
E
7
B
2
K
5
R
8
N (14 gage)
3
T
6
S
9
N (28 gage)
Type
Volts DC
Type
Volts AC
11
±100 mv
23
100 mv
12
±1 V
24
1 V
13
±5 V
25
5 V
14
±10 V
26
10 V
Summary of Contents for OMB-CHARTSCAN 1400
Page 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Page 6: ...iv ChartScan User s Manual...
Page 24: ...ChartScan User s Manual...
Page 32: ...2 8 General Information ChartScan User s Manual Notes...
Page 56: ...3 24 ChartScan User s Manual Notes...
Page 124: ...6 20 Calibration ChartScan User s Manual...
Page 126: ...A ii ChartScan User s Manual...
Page 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Page 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Page 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Page 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Page 257: ...E 2 ChartScan User s Manual Notes...
Page 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Page 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Page 271: ...ChartScan User s Manual...