6-14 Calibration
ChartScan User’s Manual
2.
Apply calibration voltage from the precision calibrator to the first channel of each card (see following
table), and verify with a digital multimeter. Requirements for both devices are specified in
Required
Equipment
.
Volts DC
Slot # 1
Slot # 2
Slot # 3
Slot # 4
Slot # 5
Slot # 6
Slot # 7
Slot # 8
-50 mV
G1,1X
G17,1X
G33,1X
G49,1X
G65,1X
G81,1X
G97,1X
G113,1X
+50 mV
G1,1X
G17,1X
G33,1X
G49,1X
G65,1X
G81,1X
G97,1X
G113,1X
-100 mV
G1,11X
G17,11X
G33,11X
G49,11X
G65,11X
G81,11X
G97,11X
G113,11X
+100 mV
G1,11X
G17,11X
G33,11X
G49,11X
G65,11X
G81,11X
G97,11X
G113,11X
-1.0 V
G1,12X
G17,12X
G33,12X
G49,12X
G65,12X
G81,12X
G97,12X
G113,12X
+1.0 V
G1,12X
G17,12X
G33,12X
G49,12X
G65,12X
G81,12X
G97,12X
G113,12X
-5.0 V
G1,13X
G17,13X
G33,13X
G49,13X
G65,13X
G81,13X
G97,13X
G113,13X
+5.0 V
G1,13X
G17,13X
G33,13X
G49,13X
G65,13X
G81,13X
G97,13X
G113,13X
-10 V
G1,14X
G17,14X
G33,14X
G49,14X
G65,14X
G81,14X
G97,14X
G113,14X
+10 V
G1,14X
G17,14X
G33,14X
G49,14X
G65,14X
G81,14X
G97,14X
G113,14X
Note:
In the slot columns, the digit after the G is the channel number and the number immediately preceding
the X assigns the voltage value, e.g., G1,1X means “Calibrate Gain for channel 1, ±50 mV.”
3.
After each calibration type
U2X
and check for a return of
U128
to confirm no errors.
Response codes and meanings applicable to calibration are as follows:
Response to
U2X
Indication for Calibration Mode
U128
No error
U129
Read Failure (EEPROM Error)
U130
Write Failure (EEPROM Error)
U132
Chksum Err (EEPROM Error)
U136
Cal Err (offset, gain, or temperature sensor out of range)
U144
Invalid password
U160
Invalid command
4.
After completing the gain calibration of all cards, type
EX
to end the calibration mode.
5.
Type
U2X
to confirm calibration mode is disabled. A return of
U000
confirms this.
Note:
You can view the calibration constants by typing an entry similar to the following:
C#1X
QC?
…for slot #1
You can monitor the calibration voltage by typing an entry similar to the following:
*B
…
to clear the buffer
C1,1X
…
to addresses the channel, this example is for slot #1, channel 1
T1,1,0,0X;@X
…
to trigger the acquisition (Trigger LED flashes); );
@X
starts acquisition
U13X
every entry of U13X results in an updated value of voltage in counts
Note that
@X
can be used to start or stop the acquisition.
Summary of Contents for OMB-CHARTSCAN 1400
Page 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Page 6: ...iv ChartScan User s Manual...
Page 24: ...ChartScan User s Manual...
Page 32: ...2 8 General Information ChartScan User s Manual Notes...
Page 56: ...3 24 ChartScan User s Manual Notes...
Page 124: ...6 20 Calibration ChartScan User s Manual...
Page 126: ...A ii ChartScan User s Manual...
Page 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Page 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Page 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Page 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Page 257: ...E 2 ChartScan User s Manual Notes...
Page 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Page 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Page 271: ...ChartScan User s Manual...