Appendix A
API Commands
ChartScan User’s Manual
A-53
S
- Set Real Time Clock
TYPE
System
EXECUTION
Immediate
SYNTAX
Stime
Set the unit’s internal real-time clock where
time
is in the standard time/date
stamp format:
HH:MM:SS:T,mm/dd/yy
where
H
is hours,
MM
is
minutes,
SS
is seconds,
T
is tenths of a second,
mm
is month,
dd
is day, and
yy
is year.
S?
Query
Stime
as defined above.
DESCRIPTION
The Set Real Time Clock (
S
) command is used to set the unit’s internal real-time clock. The
time
argument
follows the same format as the standard time/date stamp input format as specified in the
Data Format
section of
Appendix C.
The Set Real Time Clock command is used to set the battery backed internal real-time clock of the unit. In order
for the time/date stamped data to be accurate, the internal real-time clock must be set to an accurate time/date.
Once set, however, the battery-backed real-time clock will remember the current time, even when powered-off.
EXAMPLE
PRINT#1, “OUTPUT07; S14:00:00.0,4/30/93X”
‘ Current date and time
Summary of Contents for OMB-CHARTSCAN 1400
Page 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Page 6: ...iv ChartScan User s Manual...
Page 24: ...ChartScan User s Manual...
Page 32: ...2 8 General Information ChartScan User s Manual Notes...
Page 56: ...3 24 ChartScan User s Manual Notes...
Page 124: ...6 20 Calibration ChartScan User s Manual...
Page 126: ...A ii ChartScan User s Manual...
Page 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Page 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Page 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Page 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Page 257: ...E 2 ChartScan User s Manual Notes...
Page 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Page 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Page 271: ...ChartScan User s Manual...