API Commands
Appendix A
A-14
ChartScan User’s Manual
*C
- Clear Channel Configuration
TYPE
Channel
EXECUTION
Deferred
SYNTAX
*C
Clear out the channel configuration so that no channels are configured
DESCRIPTION
The Clear Channel Configuration (
*C
) command will clear out the channel configuration so that no channels are
configured. After this command has been performed, channel readings will be not be updated; therefore, no HLL
or acquisition buffer updates will be performed until a new configuration is entered via the Configure Channels (
C
)
command. This command may be used when the present configuration is no longer acceptable and you wish to
reconfigure channels from scratch.
EXAMPLE
PRINT#1,"OUTPUT07; *CX"
‘ Clear all channels configuration
PRINT#1,"OUTPUT07; C1, 1X"
‘ Configure channel 1 of T/C type J
Summary of Contents for OMB-CHARTSCAN 1400
Page 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Page 6: ...iv ChartScan User s Manual...
Page 24: ...ChartScan User s Manual...
Page 32: ...2 8 General Information ChartScan User s Manual Notes...
Page 56: ...3 24 ChartScan User s Manual Notes...
Page 124: ...6 20 Calibration ChartScan User s Manual...
Page 126: ...A ii ChartScan User s Manual...
Page 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Page 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Page 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Page 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Page 257: ...E 2 ChartScan User s Manual Notes...
Page 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Page 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Page 271: ...ChartScan User s Manual...