API Commands
Appendix A
A-12
ChartScan User’s Manual
@
- Trigger On Command
TYPE
Acquisition
EXECUTION
Deferred
SYNTAX
@
Trigger upon receipt of an
@
DESCRIPTION
The Trigger On (
@
) command will start or stop the acquisition accordingly when the
start
or
stop
arguments
of the Set Trigger Configuration (
T
) command is set to
1
(trigger on). The
T
command is used as a trigger
source and follows the same rules as other trigger sources as defined by the
T
command. The
@
command is
valid only when an acquisition is configured where either the start and/or stop event is defined as the Trigger On
command. If this command is issued when the acquisition is not configured for stop and/or start Trigger On
command, an error condition will occur.
EXAMPLE
PRINT#1, “OUTPUT07;T1,1,0,0X”
‘ Configure an acquisition with both start and stop
trigger being Trigger On command
PRINT#1, “OUTPUT07; @X”
‘ Start the acquisition
SLEEP 10
‘ Wait 10 seconds to collect data
PRINT#1, “OUTPUT07; @X”
‘ Stop the acquisition
Summary of Contents for OMB-CHARTSCAN 1400
Page 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Page 6: ...iv ChartScan User s Manual...
Page 24: ...ChartScan User s Manual...
Page 32: ...2 8 General Information ChartScan User s Manual Notes...
Page 56: ...3 24 ChartScan User s Manual Notes...
Page 124: ...6 20 Calibration ChartScan User s Manual...
Page 126: ...A ii ChartScan User s Manual...
Page 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Page 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Page 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Page 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Page 257: ...E 2 ChartScan User s Manual Notes...
Page 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Page 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Page 271: ...ChartScan User s Manual...