Appendix A
API Commands
ChartScan User’s Manual
A-49
QC?
- Query Card Data
TYPE
Calibration
EXECUTION
Immediate
SYNTAX
QC?
Read back card ID and calibration information from the card previously
selected by the Select Card (
C#
) command.
DESCRIPTION
Note:
This command is only for advanced users who perform their own calibration. It is not necessary for
normal, everyday operation.
The information returned is as follows:
•
Card Number
•
Serial Number
•
Card ID (type 16
- TC/Volts card)
(type 17 - High Volts card)
•
Gains and offsets for PGA values 0 – 7
•
Offsets and temperature sensors
•
Time and date last calibrated
Refer to the Select Card (
C#
) command for selecting chassis, card only, or card/chassis information.
EXAMPLE
PRINT#1, “OUTPUT07; C#5X”
‘ Select card #5
PRINT#1, “OUTPUT07; QC?X”
‘ Read back the card #5 information
INPUT#1 A$
PRINT#1 A$
Summary of Contents for OMB-CHARTSCAN 1400
Page 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Page 6: ...iv ChartScan User s Manual...
Page 24: ...ChartScan User s Manual...
Page 32: ...2 8 General Information ChartScan User s Manual Notes...
Page 56: ...3 24 ChartScan User s Manual Notes...
Page 124: ...6 20 Calibration ChartScan User s Manual...
Page 126: ...A ii ChartScan User s Manual...
Page 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Page 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Page 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Page 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Page 257: ...E 2 ChartScan User s Manual Notes...
Page 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Page 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Page 271: ...ChartScan User s Manual...