ChartScan User’s Manual
ChartView Software Reference 4-39
Burst Mode, Operational Issues
Acquisition Setup Dialog Box, High-Speed, Single-Channel Mode
The
Burst Mode
, also referred to as
High-speed, single-channel mode
, is selected with a radio-button located on
the lower left of the Acquisition Setup Dialog Box. The following bulleted issues apply to Burst Mode operations.
•
When in the
High-speed, single-channel (burst mode)
, all
ChartView
display functions are intentionally
disabled. Data acquisition and storage to disk are the only services that remain functional. When you
select the
High-speed, single channel mode,
a Warning message box appears, as indicated in the
following figure.
This Warning Appears Prior to Enabling the Burst Mode
•
The
Burst Mode
channel will be the only channel enabled in the instrument. This is regardless of the
ON/OFF status indicated on the
Channel and Alarm Setup
spreadsheet (located on the
Setup Channel
Types, Alarms and Acquisition Parameters
Window).
Summary of Contents for OMB-CHARTSCAN 1400
Page 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Page 6: ...iv ChartScan User s Manual...
Page 24: ...ChartScan User s Manual...
Page 32: ...2 8 General Information ChartScan User s Manual Notes...
Page 56: ...3 24 ChartScan User s Manual Notes...
Page 124: ...6 20 Calibration ChartScan User s Manual...
Page 126: ...A ii ChartScan User s Manual...
Page 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Page 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Page 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Page 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Page 257: ...E 2 ChartScan User s Manual Notes...
Page 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Page 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Page 271: ...ChartScan User s Manual...