Getting started FLUOVIEW/
Outline of LSM Observation Procedures
1-58
Page
1-2-12 Turning Power Off
1. Turn off the power supply to the units (reflected light power supply, switched power
outlet unit).
2. Turn the laser power OFF.
•
Ar laser:
Turn the key to the OFF position, then set the power switch to OFF.
(For details, refer to the instruction manual of your laser unit.)
•
Kr laser:
Turn the key to the OFF position and wait for the fan to stop automatically when
the laser unit has cooled down. Then set the power switch to OFF.
(Refer to the instruction manual of your laser unit.)
•
HeNe (Green / Red) laser:
Turn the key to the OFF position.
•
UV-Ar
laser:
1.1 Turn the key switch on the UV laser power supply to the “O-OFF” position.
1.2 Set the MAIN POWER lever on the UV laser power supply to the “O-OFF”.
Chiller is switched to auto cooling operation by interlocking.
When it is left in auto cooling, it will automatically turn off in
about 15 minutes.
While the heat exchanger is running, do not shut off the
power supply to the power outlet where the chiller is
connected.
If the power supply is turning on, the UV Ar laser tube may
be destroyed by high temperature.
After the heat exchanger has turned off, leave it for more
than an hour until the laser tube has cooled down before
turning it on again.
UV laser power supply
Key switch
MAIN POWER
TIP
Summary of Contents for FLUOVIEW FV500
Page 2: ......
Page 8: ......
Page 10: ......
Page 24: ......
Page 110: ......
Page 327: ...APPLIED OPERATIONS Viewing 3D Image 2 217 Page Fig 2 106 Panel Showing Stereo 3D Images ...
Page 396: ......
Page 402: ......
Page 414: ......
Page 420: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 458: ......
Page 460: ......
Page 462: ......
Page 468: ......
Page 469: ......