APPLIED OPERATIONS
/Image Acquisition
2-28
Page
1 Acquiring Image in Accumulation Mode (Frame mode)
1. From the page tabs on the bottom right of the [Acquire] panel, select the [Scan]
sub-panel.
Fig. 2-11 [Scan] Sub-panel
2.
In the [Filter] group box, select the [Kalman] or [Accumulate To Peak] option button.
3. When [Kalman] is selected, enter the accumulation count in the text box. And
select the [Frame Kalman] option button displayed.
The accumulation count can be set up a maximum of 63 times.
When 0 is set as the number of times of accumulation, an ordinary image
acquisition is to be performed.
[Filter] group box
Select the accumulation mode.
Two accumulation modes, [Kalman] and
[Accumulate To Peak] are available.
[Frame Kalman] option
button
Enter the accumulation
count.
TIP