Getting started FLUOVIEW/
Online Help
1-76
Page
3. Click the <
> button of the channel you want to change the pinhole diameter.
The group boxes for setting the pinhole diameters and laser ND filter values of the
channels are displayed below the [Ch] group box. (They are not displayed for
channels set for transmitted light observation.)
[PMT], [Gain] and [Offset] LED sliders
Set these values independently.
Increasing the PMT Voltage improves
the sensitivity.
Increasing the Gain brightens the image.
Increasing the Offset darkens the image.
[
Ch1], [Ch2], [Ch3], [Ch4] and [Ch5]
group boxes
Select the channels from which you want
to acquire images.
<
> button
Click to close each group box showing the
pinhole diameter and laser ND filter value.
The pinhole diameter and laser ND filter
value are valid even before the group box is
closed by clicking the <
> button.
[C.A.] and
Laser ND LED
sliders
Set these values
independently.
The laser type
used with each
channel is
displayed below
its Laser ND LED
slider.