Getting Started FLUOVIEW/
Basic Operations
1-5
Page
(3) Analyzer FV5-ANA
Polarizing plate for use in differential interference observation and polarized light
observation.
•
Set to the pushed-in position to engage the FV5-ANA in the light path for visual
transmitted light differential interference observation or transmitted polarized light
observation.
•
For laser microscope, set to the pulled-out position to disengage the FV5-ANA
from the light path. Engaging the FV5-ANA in the light path degrades image
quality.
(4) Transmitted light DIC slider
FV5-DICT
This is a prism for use in differential interference observation.
•
Engage the FV5-DICT in the light path for laser differential interference
observation or visual transmitted light differential interference observation.
Leaving the FV5-DICT engaged during laser fluorescence observation will
degrade resolution somewhat. We recommend disengaging the FV5-DICT from
the light path when simple laser fluorescence observation is required.
With transmitted light differential interference observation
using an immersion objective, set the microscope’s field
diaphragm so that it circumscribes the field of view.
Otherwise the contrast may degrade. (This applies to both
visual observation and laser differential interference
observation.)
(5) Filters
These filters are used to adjust transmitted light.
•
Be sure to disengage any filter from the light path for transmitted observation
using lasers. Leaving a filter engaged in the light path will degrade the image
quality.
NOTE
Summary of Contents for FLUOVIEW FV500
Page 2: ......
Page 8: ......
Page 10: ......
Page 24: ......
Page 110: ......
Page 327: ...APPLIED OPERATIONS Viewing 3D Image 2 217 Page Fig 2 106 Panel Showing Stereo 3D Images ...
Page 396: ......
Page 402: ......
Page 414: ......
Page 420: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 458: ......
Page 460: ......
Page 462: ......
Page 468: ......
Page 469: ......