Getting started FLUOVIEW/
Outline of LSM Observation Procedures
1-54
Page
1-2-8-9 Setting a Lower Scan Speed
The scan speed can be decreased using the scale in the [Scan Speed] group box on the
[Acquire] panel.
In general, setting a lower scan speed allows the acquired image quality to
be improved.
However, a low scan speed also lengthens the time required for image
acquisition.
When the scan speed is decreased during fluorescence observation,
the saturation of fluorescence may darken the image of certain types
of specimens. In this case, increase the scan speed and increase the
PMT Voltage or use accumulation in scanning.
1-2-8-10 Stopping Repeated Scanning
After the brightness and gain have been adjusted, select the <STOP SCAN> button in
the [Acquire] panel to stop scanning temporarily.
[Scan Speed] group box
Set the scan speed by clicking a
point on the scale line.
TIP
NOTE