Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
1-31
Page
1-2-4 Setting the LSM Light Path
1-2-4-1 Combination with Upright Microscope (BX)
1. From the page tabs on the bottom right of the [Acquire] panel, select the [Optics]
sub-panel.
Fig. 1-6 [Optics] Sub-panel
2. Select the <LSM> button in the [Light Path] group box.
The <LSM> button looks pushed in to indicate that it is selected.
(When scanning is started while the <BI> button is selected, the LSM light path is
selected automatically. It is switched back to the visual observation automatically
when scanning completes.)
3.
Push the hand switch button to (8) set
to be displayed in the cube display
window (1) on the reflected light fluorescence vertical illuminator
4. When the U-MDICT3 analyzer (2) is in use, disengage it from the light path by
setting the switch to the pulled-out position.