ML41

Biological microscope

User manual

Website

www.m-shot.com

Tel

0086-20-38262481

Summary of Contents for ML41 Series

Page 1: ...Guangzhou Micro shot Technology Co Ltd ML41 Biological microscope User manual Website www m shot com Tel 0086 20 38262481 ...

Page 2: ...Biological microscope ML41 ...

Page 3: ...ry or damage to this unit Please contact the supplier if any fault occurs 3 Input voltage Check if the input voltage is consistent with your local voltage supply If not do not operate this unit and contact the supplier Improper input voltage may cause a short circuit or fire thereby causes damage to this unit 4 Use specific bulb fuse and power cord Use of an improper bulb fuse or power cord may ca...

Page 4: ... temperatures dampness or dust for a prolonged period of time otherwise mist or mold may develop or dust may deposit on the lens thus cause damage to this unit and shortening its life 9 Installation of bulb Do not touch the glass surface of the bulb directly with bare hands When mounting the bulb wear gloves or wrap it with cotton material Wipe off any dirt on the surface of the bulb with a clean ...

Page 5: ...ML41 biological microscope user manual Website www m shot com V sale mshot com ...

Page 6: ...user manual sales mshot com VI Website www m shot com Content Ⅰ Attention 7 Ⅱ Product characteristic 9 1 Name of each part 9 2 Installation of complete machine 11 Ⅲ Instructions 13 Ⅳ Specifications 17 Ⅴ Common Troubleshooting 19 ...

Page 7: ...The power supply socket shall be provided with grounding wire 3 Handling and placement 1 Turn off the power switch before handling the instrument 2 Working environment requirements Room temperature 0 C 40 C Maximum relative humidity 85 3 Avoid placing the instrument under direct sunlight or room light The bright environment will affect the imaging quality of the specimen 4 Avoid placing the micros...

Page 8: ...ite www m shot com the left and right coarse focusing handwheel to rotate in reverse to each other otherwise it will cause serious loss of focusing accuracy or even failure 2 Do not touch the surface of optical parts by hand otherwise the imaging will be blurred ...

Page 9: ...ical microscope user manual Website www mshot com 9 sales mshot com Ⅱ Product characteristic 1 Name of each part Name of each component of ML41 series laboratory biological microscope Light color temperature wheel ...

Page 10: ...ML41biological microscope user manual Sales mshot com 10 Website www m shot com ...

Page 11: ...w to install each component and the number indicates the sequence of installation before installation please make sure that all parts of the microscope are free of dust dirt and other foreign matters affecting the installation During installation be careful not to scratch or scratch any part or touch the glass surface Observation head Eyepiece Power line Objective Stage Condenser ...

Page 12: ... scale facing forward and align the set screw on the back of the condenser with the slot on the condenser holder and push the condenser into the innermost part of the condenser holder Picture 4 5 Tighten the condenser fixing screw then turn the condenser lifting hand wheel to raise the condenser to the highest position 2 2 3 Install the objectives Turn the coarse hand wheel to lower the stage and ...

Page 13: ...er cord is easily damaged when bent and twisted please do not use excessive force Please use the power cord provided by us and choose another suitable power cord if it is lost or damaged The power socket should have a ground terminal to ensure that the instrument can be safely grounded Ⅲ Instructions 3 1 Lighting adjustment 1 Turn on the power put the switch in the 1 state and the bulb will light ...

Page 14: ...e direction shown in the picture The coarse adjustment limit hand wheel can prevent the slice and the objective lens from colliding during focusing Coarse adjustment limit hand wheel has no effect on micro hand wheel 3 Turn the micro hand wheel to clearly observe the details of the specimen Picture 11 When using the 4X and 10X objectives the aperture diaphragm and field diaphragm are opened to the...

Page 15: ...Figure 13 the indication point above the left eyepiece holder points to the scale on the pupil distance sign and the corresponding value is the current observer s pupil distance Please remember your pupil distance value for your next use 3 7 Adjusting the stage During observation the specimen can be moved to the x axis and y axis respectively by rotating the x axis and y axis moving knobs of the s...

Page 16: ...mage circumscribes to the field 3 9 Field diaphragm The field aperture limits the diameter of the light beam entering the condenser so as to block the excess light at the periphery and enhance the image response When the imaging of the field aperture is just at the outer edge of the field of view the objective lens can play the most superior performance and get the clearest specimen image 3 10 Ape...

Page 17: ... aperture it should be round and bright b When eliminating the bubbles slightly turn the nose piece and wave the oil immersion objective back and forth for several times 4 After finished using the oil immersion objective carefully wipe the front lens with clean gauze and or lens wipes dipped in Ethanol ethyl ether mixture mixed with 3 7 or xylene Do not shift other objectives in before the oil imm...

Page 18: ...17 Attention The refraction index of the immersion oil is 1 515 4 2 parameters for eyepieces Categories Magnification Field of view mm High vision plan eyepiece 10 Φ23 4 3 Total magnification total 4 10 40 100 10 40 100 400 1000 4 4 Condenser Swing Out Achromatic Condenser NA1 2 0 22 4 5 Coaxial coarse and fine focus adjustment mechanism with coarse adjustment upper limit and elastic adjustment de...

Page 19: ...g tips If you still can t find the cause of the fault please find an experienced professional for repair or contact the manufacturer directly Fault Cause Troubleshooting method Specimen imaging is not clear Put the specimen upside down Reposition the specimen Inappropriate coverslip thickness Use 0 17mm thick coverslips Dirt on the lens wipe clean Oil immersion objective is not oiled Add cedar oil...

Page 20: ...condenser is not aligned in the center Adjust the condenser center The bulb brightness is turned to maximum but the field of view lighting is not bright enough The field diaphragm is not wide enough Open the field diaphragm wider The condenser is not aligned in the center Adjust the condenser center Condenser position is too low Raise the condenser Aperture diaphragm opened too small Appropriately...

Page 21: ...ng and dizziness The left eyepiece tube diopter is not adjusted properly Readjust the left eyepiece tube diopter Coarse handwheel rotates very tightly The elastic adjustment ring is locked too tightly Lossen appropriately The stage descends automatically first and cannot stop on the focal plane during observation The elastic adjustment ring is too loose Lock appropriately Coarse handwheel cannot r...

Page 22: ...ll amount of pure ethanol ether mixture mixed in a ratio of 3 7 to gently wipe off 6 3 Do not use organic solvents to wipe the non optical surfaces of the microscope Use gauze dipped in clean water to wipe these surfaces For stubborn stains neutral detergents can be used to clean them Both ethanol and ether are extremely flammable Be careful not to place these chemicals near working light sources ...

Page 23: ...moisture proof tank Pay attention to put the original dust cap and converter plug back into the eyepiece tube and on the converter after removing the eyepiece and objective lens Product design is continually being improved but these improvements may not be described in this manual ...