6.3.3
Virtual Port Software Installation
1.
Insert the supplied CD into the computer’s
CD drive.
Different versions of Windows have slightly
different steps to load the driver that is on
the CD. In all versions the New Hardware
Wizard guides you through the required
steps to select the driver that is located on
the CD.
2. After clicking Finish, the virtual port should
be ready for use.
Windows typically adds the virtual port in
sequence after the highest number COM
port. For example, on PC’s equipped with
up to 4 COM ports, the virtual port will be
COM5.
When using the USB interface with
programs that limit the number of COM
port designations (e.g. Ohaus
MassTracker allows only COM1, 2, 3, & 4),
it may be necessary to assign one of these
port numbers to the new virtual port.
Example of Windows XP Hardware Wizard
This can be done in the Port Settings of the Device Manager utility, found in the Windows Control Panel.
6.4
USB Host
The USB Host can be used to connect a keyboard, barcode scanner and USB flash drive to the Ranger 7000.
6.5
Printout Format
Printout string for g, kg, lb, oz units:
Field
Weight Space Unit Space Stability Space G/N Space Message Term.Char(s)
Length 9
1
3
1
1
1
1
1
5
2
The printout string has a fixed length of 23 characters.
Each Space field is a delimiting space used to separate the other fields.
The Weight field is 9 right justified characters. If the value is negative, the '
‐
' character is printed at the
immediate left of the most significant digit.
The Unit/Mode field is 3 left justified characters.
The Stability field is 1 character.
The G/N field is 1 character. 'G' is printed for a gross weight. 'N' is printed for a net weight.
The Message field is 5 left justified characters.
Note:
The Termination Characters Carriage Return and Line Feed are appended to the printout.
Printout string for the lb:oz unit
Field
Weight1 Space Unit1 Space Weight2 Space Unit2 Space Stability Space G/N Space Message Term.Char(s)
Length 4
1
2
1
7
1
2
1
1
1
1
1
5
2
The printout string has a fixed length of 28 characters.
Each Space field is a delimiting space used to separate the other fields.
The Weight1 field is 4 right justified characters. If the value is negative, the '
‐
' character is located at the
immediate left of the most significant digit.
The Unit1 field is 2 left justified characters.
The Weight2 field is 7 right justified characters.
The Unit2 field is 2 left justified characters.
The Stability field is 1 character. A space is printed if the weight value is stable. A '?' is printed if the weight
value is not stable.
The G/N field is 1 character. 'G' is printed for a gross weight. 'N' is printed for a net weight.
The Message field is 5 left justified characters.
Note:
The Termination Characters Carriage Return and Line Feed are appended to the printout.