RANGER
®
7000 SCALES
EN-41
5.2
Main Menu
The Main menu selections are illustrated below.
Calibration
Setup
Read Out
Application
Modes
Weighing
Units
GLP and
GMP Data
Communication
User Profile
Alibi
Memory
Event Counter
Maintenance
5.3
Calibration
R71MD models offer three calibration methods:
Zero Calibration, Span Calibration and Linearity Calibration.
R71MHD models offer 5 calibration methods:
Zero Calibration, Span Calibration, Linearity Calibration, Internal Calibration and
Automatic Calibration.
Do not disturb the scale during any calibration.
5.3.1
Calibration sub-menu
R71MD models:
Zero Calibration
Span
Calibration
Linearity Calibration
GEO
R71MHD models:
Zero Calibration
Span
Calibration
Linearity
Calibration
Internal
Calibration
Automatic
Calibration*
5.3.2
Zero Calibration
Use this calibration method to adjust the zero calibration point, without affecting the span or linearity
calibration.
5.3.3
Span Calibration
Span calibration uses two calibration points, one at zero load and the other can be chosen by the user by
using the numerical keypad.
With the scale turned ON and no load on the pan, start Span Calibration to initiate the
procedure. Additional calibration values to be used are shown on the display. The best
accuracy is achieved using the mass closest to the full span value.