9.5.5 I/O Menu “Sensor-Info”
This menu is to indicate various information regarding the sensor and is mainly
for service purposes.
1
Firmware version of sensor
2
Creation date of sensor firmware
3
Quality of velocity measurement
4
Calculated average flow velocity
5
Height measured by pressure sensor
6
Measured medium temperature
7
Velocity of sound resulting from medium temperature
8
Amplification mode of sensor
9
Amplification value of sensor
Fig. 9-59
Status of sensors and velocity evaluation
9.5.6 I/O-Menu „v-Histogram”
The frequency histogram indicates the spreading of the investigated Doppler
frequency. Each bar (peak) represents a frequency group.
This is particularly important to assess and to choose measurement places as
well as to find a place for sensor installation.
1 = Quality of velocity measurement
2 = frequency group (peak)
3= Measurement frequency
4= Currently measured flow velocity
5= invalid values
Fig. 9-60
Distribution of frequency groups
The measurement quality (0-100 %) indicates the relation between the evaluat-
ed Doppler frequency and the entire frequency spectrum measured. The higher
the quality, the more reliable the indicated flow velocity reading. There are no
limit values for the quality (Q) since the shape of the frequency distribution has
to be considered additionally. This shape is more important for hydraulic as-
sessment than „q“.