Instruction Manual
OCM F
Page 66
OCM F - Rev. 04 as of 18.02.2014
®
Fig. 9-52
New program start OCM F
Fig. 9-53
Choose language
The OCM F will now overwrite the flash memory restarting the program. Before
restart the unit briefly indicates unit ID (see Fig. 9-52). Subsequently the desired
language can be selected (see Fig. 9-53) and the unit will return to factory de-
fault condition.
Service mode
Additional system setting options are going to be revealed as soon as a spe-
cial code has been entered. This parameter is reserved for NIVUS service
personnel as the settings require comprehensive expert knowledge and are
not needed for common applications.
Damping
Allows to adjust indication and analog output damping between 20 and 200
seconds. This means that if the calculated volume jumps from 0 to 100 % the
system will need the time set here to indicate and to output this jump.
Stability
It is recommended to increase stability in case of irregular measurement drops
due to poor hydraulic conditions..