Instruction Manual
OCM F
OCM F - Rev. 04 as of 18.02.2014
Page 31
Fig. 7-16
Overvoltage protection compact Doppler active sensor
Fig. 7-17
Overvoltage protection compact Doppler active sensor (Ex)
When using the sensors in Ex areas the electrical values of overvoltage pro-
tection elements as well as capacities and inductance of the KDA sensor ca-
ble shall be considered additionally!
The NIVUS cable lengths below are allowed for use in Ex areas:
- unilateral overvoltage protection: 135 m
- double-sided overvoltage protection: 120 m
Please observe the non-reversed connection (p-side to transmitter) as well as
a correct, straight wiring supply.
Ground (earth) must lead to the unprotected side. Incorrect wiring will deacti-
vate the overvoltage protection function!