OPERATING PROCEDURE: Printing Measured Values
3.8
Printing Measured Values
Three printing settings (“PRINT1” to “PRINT3”) are available to customize the printing contents. Set
the parameters as necessary or desired.
PRINT1 (Print setting 1)” (page 92)
PRINT2 (Print setting 2)” (page 93)
,
1
When measurement is complete, print the measured values.
The following two methods are available for printing by infrared communication.
●
Place the main body on the station, then press
.
Placing the main body on the station offers more
stable communication.
●
Aim the main body toward the IR receptor B of the
station, then press
.
The maximum communication distance
between the main body and station is 1 m.
Main body re-
ceptacle
Within 1 m
IR receptor B
• When the
parameter is set to “AUTO”, printing starts automatically without
having to press
.
• When the
parameter is set to “IR (infrared communication)”, press
. “AUTO” cannot be selected for the