PARAMETER SETTINGS AND MAINTENANCE: Parameter Settings
●
PRINT1 (Print setting 1)
PRINT1 parameter screen
Parameter option
Setting contents
31. PRINT
(Printing)
MANUAL / AUTO / NO
Sets the printing operation from among MANUAL, AUTO, and NO.
•
MANUAL: Pressing the print button starts printing.
•
AUTO: Printing starts automatically when measurement is complete.
•
NO: Printing does not occur.
* When the
“AUTO” cannot be selected.
32. OUTPUT W/PRINT
(Data output during printing)
YES / NO
Sets whether to output data during printing.
When set to “YES”:
•
Data is output to the destination specified by the
parameter.
•
When the
parameter is set to “YES”, the
data is output to the refractor.
33. MEMORY PRINT
(Printing memory data)
PRINT / OUTPUT /BOTH
Sets the output type of memory data.
•
PRINT: Only printing occurs.
•
OUTPUT: Only data output occurs.
•
BOTH: Both printing and data output occur.
•
Data is output to the destination specified by the
. When the
parameter is set to
“YES” with only one data being selected, the data is output to the refractor.
34. ECONO. PRINT
(Economical printing)
YES / NO
Sets whether to print the data with reduced line-spacing to save printer
paper.
35. PRINT&CLEAR
(Erasing of data after printing)
YES / NO
Sets whether to delete the measured data when the next measurement is
performed after printing.
* When the
parameter is set to “IR (infrared
communication)”, the data is not automatically deleted.