120
SPECIFICATIONS AND TECHNICAL INFORMATION: EMC (Electromagnetic Compatibility)
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure
that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in ISM and
amateur radio
bands between 0.15
MHz and 80 MHz
80% AM at 1 kHz
3 V
0.15 MHz to 80 MHz
6 V in ISM and
amateur radio
bands between 0.15
MHz and 80 MHz
80% AM at 1 kHz
Portable and mobile RF communications equipment should be used
no closer than 30 cm to any part of the device, including cables.
a
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
Proximity field from
RF wireless
communications
equipment
IEC 61000-4-3
See “Test
specifications
for enclosure port
immunity to RF
wireless
communications
equipment”.
See “Test
specifications
for enclosure port
immunity to RF
wireless
communications
equipment”.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
device.
Test specifications for enclosure port immunity to RF wireless communications equipment
Test
frequency
(MHz)
Band
a
(MHz)
Service
a
Maximum power
(W)
Distance
(m)
Immunity test level
(V/m)
385
380 - 390
TETRA 400
1.8
0.3
27
450
430 - 470
GMRS 460, FRS 460
2
0.3
28
710
704 - 787
LTE Band 13, 17
0.2
0.3
9
745
780
810
800 - 960
GSM 800/900, TETRA 800,
iDEN 820, CDMA 850, LTE Band 5
2
0.3
28
870
930
1,720
1,700 - 1,990
GSM 1800, CDMA 1900, GSM 1900,
DECT; LTE Band 1, 3, 4, 25: UMTS
2
0.3
28
1,845
1,970
2,450
2,400 - 2,570
Bluetooth, WLAN, 802.11: b/g/n,
RFID 2450, LTE Band 7
2
0.3
28
5,240
5,100 - 5,800
WLAN 802.11
a/n
0.2
0.3
9
5,500
5,785
NOTE: If necessary to achieve the immunity test level, the distance between the transmitting antenna and the ME equipment or ME
system may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.