112
PARAMETER SETTINGS AND MAINTENANCE: Maintenance
3
Check if the window is cleaned using a penlight. If soiled areas remain, clean the
window again with new cleaning paper.
O
Cleaning the printer
After repeated usage, the paper slot of the auto cutter of the printer may become soiled with powdery
paper residue. If paper residue settles, malfunction of the auto cutter may result. Clean it periodically.
1
Press
to open the printer cover and
remove the printer paper.
2
Apply the nozzle of a vacuum cleaner to the
auto cutter to remove paper residue.
Never blow off paper residue with a blower. If
paper residue settles on the internal working
structure, malfunction may result.
3
Supply the printer paper as it was.
4.2.5
Consumable list
• When the
parameter is set to “YES” or “DAY”, the measuring
window is checked for cleanliness at device start-up.
Printer paper
Part name
Part number
Remarks
Printer paper
8062000001
Width 58 mm, Length 25 m
Battery pack
30621-9202
7.2 V, 1800 mAh