![Nidek Medical HandyRef-K Operator'S Manual Download Page 99](http://html1.mh-extra.com/html/nidek-medical/handyref-k/handyref-k_operators-manual_1688062099.webp)
93
PARAMETER SETTINGS AND MAINTENANCE: Parameter Settings
●
PRINT2 (Print setting 2)
36. PRINT DENSITY
(Density of printed text)
LOW / MEDIUM / HIGH
Sets the density of the printed text from among LOW, MEDIUM, and HIGH.
37. PATIENT NO.
(Printing of patient number)
YES / NO
Sets whether to print the patient number.
38. SET PATIENT NO.
(Setting of patient ID)
0001 to 9999
Sets the patient number in the range from 0001 to 9999.
Move the highlight to SET PATIENT NO. and press [
→
] or [
←
] to change the
patient number.
•
[
→
] button: Advances the patient number.
•
[
←
] button: Reduces the patient number.
39. SUMMARY
(Display of summary screen)
YES / NO
When “YES” is selected, finishing measurement for both eyes or pressing
the print button displays the summary screen before the data is printed out.
PRINT2 parameter screen
Parameter option
Setting contents
51. AR PRINT
(Print format of AR data)
ALL / SHORT
Selects the print format of AR data.
•
ALL: All data and median values are printed.
•
SHORT: Only median values are printed (when median values have not
been obtained, the latest values are printed).
52. KM PRINT
(Print format of KM data)
ALL / ALL (KM) / SHORT
Selects the print format of KM data.
•
ALL: All data and median values are printed.
•
ALL(KM): In KM measurement mode, all data and median values are
printed. In other modes, only median values are printed.
•
SHORT: Only median values are printed.
53. CONF. INDEX
(Printing of confidence index)
YES / NO
Sets whether to print the confidence index.