![Nidek Medical HandyRef-K Operator'S Manual Download Page 124](http://html1.mh-extra.com/html/nidek-medical/handyref-k/handyref-k_operators-manual_1688062124.webp)
118
SPECIFICATIONS AND TECHNICAL INFORMATION: EMC (Electromagnetic Compatibility)
O
Specified cable
O
Specified multimedia equipment
•
Network switch: Complied with CISPR 32 Class B
•
Computer: Complied with CISPR 32 Class B
O
Essential performance
•
Objective refraction function
•
Keratometry measurement function
Part name
Connector
shielded
Cable shielded
Ferrite core
Length (m)
Power cord
No
No
No
2.5
Connection cable
Yes
Yes
Yes
2.0
LAN cable
Yes
Yes
No
5.0
Communication cable
Yes
Yes
Yes
15.0
USB cable
Yes
Yes
No
2.0