PARAMETER SETTINGS AND MAINTENANCE: Parameter Settings
4.1.4
Setting station parameters
●
When setting station parameters
On the PARAMETER SETTING screen, select “SET
STATION” with
or
button and press
.
The station parameter settings set by the main body
are applied to the station.
While the settings are being applied, the message
“SETTING OF STATION” is displayed. When com-
plete, “SETTING FINISHED” is displayed.
After the parameter settings are changed, press
to return to the Main screen.
●
Before performing “SET STATION” parameter
Before performing
“SET STATION”
parameter, be
sure to connect the main body to the station with the
connection cable.
If the cable is not connected, transfer of the parame-
ter settings to the station will fail, and the screen
shown to the right appears.
1)
Confirm that the cable is securely connected to
the main body and station and that the stations
is turned on.
2)
On the screen shown to the right, select “YES”
with
or
button and press
.
After setting any station-related parameter, should the PARAMETER SETTING screen be exited with-
out performing
“SET STATION”,
the parameter settings will still be automatically transferred to the
station.
In this case as well, make sure that the main body and station are connected with the connection
cable. If the cable is not connected, transfer of the parameter settings to the station will fail, and the
screen shown to the right appears.