INTRODUCTION: Before First Use
2.4.8
Powering the station by battery pack
Normally, the station operates by an AC power supply. However, the station can be powered by the
battery pack by the following procedure.
This is a supplemental function. In principle, it is recommended to connect the station to an AC power
supply.
1
Turn off the power switch of the station.
With its terminals facing the direction indicated by the arrow, insert a charged battery pack into the
battery slot.
2
Press .
The pilot lamp lights up. The station is now powered by the battery pack. To turn off the device, hold
down the button for more than a second.
• When the station is powered by the battery pack, the station will be automatically turned off by the
auto shutdown function if there is no operation or data transmission to the station for 5 minutes. Five
beeps sound before the station turns off.
When the station is turned off (the pilot lamp is off) before use, press
to turn on the power.
• When the battery charge is low, three beeps sound and the pilot lamp starts blinking. Power the sta-
tion from an AC power supply and charge the battery pack.
• When the station is powered by the battery pack, turning on the power switch of the station automat-
ically changes to the AC power. At the same time, the station battery charge indicator lights up indi-
cating that the battery pack is being charged.
Terminals
Station battery charge
indicator
Battery pack